INFINITESIMAL FIELD SIZE HVL OF ISO 4037 NARROW SPECTRUM X-RAYS: SIMULATED RESULTS

Half-value layer (HVL) measurement is the method most often used to validate an X-ray reference field as defined in ISO 4037. This work presents Particle and Heavy Ion Transport code System (PHITS) simulated values of HVL and their standard deviations for an infinitesimal radiation field size. The H...

Täydet tiedot

Tallennettuna:
Bibliografiset tiedot
Päätekijä: Le, Ngoc Thiem
Aineistotyyppi: Artikkeli
Kieli:English
Julkaistu: Trường Đại học Đà Lạt 2024
Aiheet:
Linkit:https://scholar.dlu.edu.vn/thuvienso/handle/DLU123456789/256920
https://tckh.dlu.edu.vn/index.php/tckhdhdl/article/view/1325
Tagit: Lisää tagi
Ei tageja, Lisää ensimmäinen tagi!
Thư viện lưu trữ: Thư viện Trường Đại học Đà Lạt
Kuvaus
Yhteenveto:Half-value layer (HVL) measurement is the method most often used to validate an X-ray reference field as defined in ISO 4037. This work presents Particle and Heavy Ion Transport code System (PHITS) simulated values of HVL and their standard deviations for an infinitesimal radiation field size. The HVL values, calculated from simulated values of air kerma at a detector position for various radiation field sizes, were extrapolated to an infinitesimal field size for different ISO 4037 narrow spectrum X-ray beam qualities at two distances from an X-ray source.