INFINITESIMAL FIELD SIZE HVL OF ISO 4037 NARROW SPECTRUM X-RAYS: SIMULATED RESULTS

Half-value layer (HVL) measurement is the method most often used to validate an X-ray reference field as defined in ISO 4037. This work presents Particle and Heavy Ion Transport code System (PHITS) simulated values of HVL and their standard deviations for an infinitesimal radiation field size. The H...

Full beskrivning

Sparad:
Bibliografiska uppgifter
Huvudupphovsman: Le, Ngoc Thiem
Materialtyp: Artikel
Språk:English
Publicerad: Trường Đại học Đà Lạt 2024
Ämnen:
Länkar:https://scholar.dlu.edu.vn/thuvienso/handle/DLU123456789/256920
https://tckh.dlu.edu.vn/index.php/tckhdhdl/article/view/1325
Taggar: Lägg till en tagg
Inga taggar, Lägg till första taggen!
Thư viện lưu trữ: Thư viện Trường Đại học Đà Lạt
Beskrivning
Sammanfattning:Half-value layer (HVL) measurement is the method most often used to validate an X-ray reference field as defined in ISO 4037. This work presents Particle and Heavy Ion Transport code System (PHITS) simulated values of HVL and their standard deviations for an infinitesimal radiation field size. The HVL values, calculated from simulated values of air kerma at a detector position for various radiation field sizes, were extrapolated to an infinitesimal field size for different ISO 4037 narrow spectrum X-ray beam qualities at two distances from an X-ray source.